Brookfield Bath, Model TC-101 User Manual

Page 9

Advertising
background image

10

3.2

To Change Pump Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

(also in Section 3.3, System Set Up, FCN 6 - Function Tuning)

1. Press , The pump speed menu will appear first.

2. Select the pump speed number, then press

3. See Section 2.5 Pump for pump specifications.

3.3

System Set Up

Each numerical keypad button permits access to all control parameters.

Press and hold and then select the desired operating parameter.

— Adjust Display Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Adjust the display’s contrast to compensate for ambient light and viewing angle.

Press to adjust.

— Set Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Select temperature high & low limits. If fluid temperature exceeds these limits, the unit

will shut off and the alarm will activate.

Maximum high limit value that can be entered is 203°C,

Maximum low limit is -53°C (minus sign must be entered first).

— Change Readout - °C, °F, °U . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Select desired temperature scale pressing:

°C = °F = °U = Display will momentarily read " CONVERTING"

To set °U, user scale, refer to subsection FCN 7 -°U Cal.

— Change Resolution - 0.1°, 0.01°, 0.001° . . . . . . . . . . . . . . . . . . . . . .

Change desired resolution value pressing:

0.1°= 0.01°= 0.001°=

While readout can display three decimal places, actual resolution is only .005°C.

— External Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Used to select temperature control via internal sensor or external remote probe
(optional).

By selecting 2 = EXT, the external sensor mode, the display will read . . . . . . . . . . . . .

SELECT PUMP SPEED
HIT 1 THROUGH 5

CONTRAST ADJUST
PRESS UP OR DN

HI LIMIT

=

203.00°C

LO LIMIT

=

-53.00°C

The five selectable speeds

FULL =

HIGH =

MEDIUM =

LOW =

SLOW =

1 = °C 2 = °F 3 = °U
4 = 1° 5 = .01° 6 = .001°

1 = INTERNAL SENSOR
2 = EXT 3 = CAL EXT

MAX SETPOINT
DIFFERENTIAL

= 10°C

1 = °C 2 = °F 3 = °U
4 = 1° 5 = .01° 6 = .001°