6 system generated reports, 6 operation 26 – BUCHI Melting Point M-565 User Manual

Page 26

Advertising
background image

6 Operation

26

MeltingPoint Monitor 1.1 Operation Manual, Version B

6.6

System generated reports

6.6.1

Calibration report

After a successful system calibration, the software will automatically generate a report in PDF format
if the PC is linked. In case the link has not been established, the report is stored in the temporary
memory and can be received via Tools > Get Latest Calibration.

NOTE

During calibration, viewing existing results or changing methods stored on the computer is

possible.
The calibration result stays available on the Melting Point M-565 until it is switched-off or a new

determination is started!

Temp. gradient

0.5 °C/min

Standard deviation limit

0.2 °C

No. of required samples

6

4-Nitrotoluene

Lot: 1234567890

Automatic:

Left

Center

Right

1:

50.1 °C

52.5 °C

52.3 °C

2:

54.9 °C

52.5 °C

3:

52.6 °C

53.7 °C

4:

52.8 °C

52.4 °C

56.0 °C

Average

52.5 °C

Standard deviation

0.17 °C

Reference temperature

52.0 °C

Diphenylacetic Acid

Lot: Subst. 2

Automatic:

Left

Center

Right

1:

148.1 °C

148.2 °C

148.3 °C

2:

148.9 °C

148.5 °C

3:

148.6 °C

148.7 °C

4:

148.8 °C

148.4 °C

148.0 °C

Average

148.3 °C

Standard deviation

0.19 °C

Reference temperature

148.0 °C

Caffeine

Lot: Coffein

Automatic:

Left

Center

Right

1:

237.1 °C

237.2 °C

237.3 °C

2:

237.9 °C

237.5 °C

3:

237.6 °C

237.7 °C

4:

237.8 °C

237.4 °C

237.0 °C

Average

237.3 °C

Standard deviation

0.19 °C

Reference temperature

237.0 °C

Potassium Nitrate

Lot: Kalibriersubstanz 4

Automatic:

Left

Center

Right

1:

335.1 °C

335.2 °C

335.3 °C

2:

335.9 °C

335.5 °C

3:

335.6 °C

335.7 °C

4:

335.5 °C

335.4 °C

335.0 °C

Average

335.5 °C

Standard deviation

0.14 °C

Reference temperature

335.0 °C

Last calibration

1/16/2009

Serial number

12.12.12.1

Last verification

10/30/2008

Firmware version

0.39.1

Date

1/16/2009

Time

9:54 AM

M-565

CompanyName

Calibration

MeltingPoint Monitor V1.0

Date, signature

Calibration1162009.pdf

1 / 1

Advertising