Operation Manual, Version B

4 Description of function

17

A plot of log I versus log m provides a linear response. The values of the constants (k and b) are
dependent on a variety of parameters (e.g. the temperature and the nature of the mobile phase).

An inlet to provide additional gas is located immediately before the detector chamber to provide
a concentric shield for the carrier gas. This serves to eliminate diffusion of the carrier gas and
eliminate contamination of the detection chamber.

4.3

Eluent flow splitting

The C-650 detector is designed to work with flow rates from 0.1 to 5 mL/min. In preparative chro-
matography applications, the flow rate is mostly higher than 5 mL/min. The BUCHI preparative
chromatography systems can deliver flow rates up to 250 mL/min. For preparative applications,
the eluent flow rate coming out of the chromatography column must be split into 2 different flow
rates:

•

One part (usually about 1 mL/min) is directed to the C-650 and this part of the sample is
destructed by the evaporation and diffraction process.

•

The other part is directed to the fraction collector to collect the detected compounds.

The working principle is shown below:

Pumping system

Co

lu

m

n

UV

Detector

(optional)

ELSD

Fraction

Collector

Flow splitter

≤ 250mL/min

≤ 5mL/min

Fig 4.6: Working principle of the flow splitting