5 putting into operation, 1 installation site, Putting into operation – BUCHI ELS Detector C-650 User Manual

Page 18: Installation site

Advertising
background image

C-650

Operation Manual, Version B

5 Putting into operation

18

5

Putting into operation

This chapter describes how the device is installed and gives instructions on initial startup.

NOTE
Inspect the device for damages during unpacking. If necessary, prepare a status report immedi-
ately to inform the postal company, railway company or transportation company.
Keep the original packaging for future transportation.

5.1

Installation site

Put the device on a stable, horizontal surface. Consider the maximum product dimensions and
weight. Obtain the environmental conditions as described in section 3.2 Technical data overview.

Installation prerequisites:

•

Make sure, the device is not subject to drafts or significant temperature changes.

•

Do not place it near air conditioning vents, windows, ovens, etc.

•

Make sure that access to the power supply to disconnect the device must be kept at any
time.

•

Make sure to leave a distance of at least 20 cm between the device and another object or
wall.

•

Place the detector close to the outlet of the column to minimize extra column band broaden-
ing which would reduce the resolution of the choromatographic separation.

•

WARNING

Death or serious injuries by use in explosive environments.

• Do not operate the device in explosive environments

• Do not operate the device with explosive gas mixtures

• Before operation, check all gas connections for correct installation

• Directly withdraw released gases and gaseous substances by sufficient ventilation

CAUTION

Risk of minor or moderate injury by heavy weight of the device.

• Consult a second person to lift the device

• Do not drop the device

• Place the device on a stable, even and vibration-free surface

• Keep limbs out of crushing zone

• To transport the device, place your hands under the device

NOTICE

Risk of device damage by liquids or mechanical shocks.

• Do not spill liquids over the device or its components

• Do not move the device when it is loaded with sample liquid

• Keep external vibrations away from the device

• Safely attach the device to the bench in earthquake prone regions

NOTE
As a destructive detector, the C-650 should be the last one in the flow path or can be used with
a splitter.