6 Operation

37

K-439 Operation Manual, Version A

!

WARNING

Death or serious poisoning by contact or incorporation of harmful substances at use.

Before operation, check the instrument for correct assembling

•

Close unused suction module positions with glass caps

•

Rinse suction modules, hoses and connectors before use

•

Before operation inspect sealings and tubes for good condition and tightness

•

Exchange worn out or defective parts immediately

•

Only operate the instrument in a fume hood

•

Directly withdraw released gases and gaseous substances by a Scrubber or a water jet pump

•

Dispose of condensate in tubings and glassware safely after use

•

!

CAUTION

Risk of minor or moderate cuts when handling damaged glass parts.

Handle glass parts with care

•

Do not heat up empty sample tubes in the device

•

Visually inspect every glass part before mounting

•

Exchange damaged glass parts immediately

•

Do not touch cracks or bits of broken glass with bare hands

•

6 .7 .1

Kjeldahl digestion —preparing sample tubes

During the preparation process, handling with strong acid and other dangerous substances is inevi-
table. Thus, all steps must be performed under safe laboratory conditions!

Prerequisites

All sample tubes must be clean and free of damage.

•

To achieve a reliable and reproducible digestion result the quality of sample material and additives

•

(e.g. acid and catalyst) must be free of impurities and of adequate amount.

Filling of a sample tube

Take the sample tubes sequentially and weigh in sample material.

•

Add calatyst to the samples (e.g. Kjeldahl tablets, order no. 028765).

Carefully(!)

•

add sulfuric acid (

≈ 98 %) of sufficient amount for the samples.

➡ The sample tubes can now be installed in the rack.
Note

Note down the sample weight and position to keep the probe processing and its analysis repro-

•

ducible. All racks are marked on top with a "1" for the first sample position.
The rack(s) can be used as a temporary storage for empty tubes during the preparation process.

•

When adding the sulfuric acid, rotate the tube to wash down any sticky sample material to the

•

bottom and to coat the inside of the tube.