6 Operation

41

K-439 Operation Manual, Version A

If applicable, follow the previous steps to install a suction module on a second rack.

•

Switch on the Scrubber or water jet pump and activate the fume hood ventilation.

•

➡The rack(s) can now safely be processed in the SpeedDigester.
Note

Clean all glass parts before mounting to avoid cross contamination.

•

When a water jet pump is used, hazardous vapours can escape into the environmental air if the

•

suction is inadequate.
If a Scrubber without condensation step is used, the condensate that accrues must be absorbed

•

by means of a Woulff bottle between the SpeedDigester K-439 and the Scrubber.
When operating with 3rd party sample tubes no insulation caps and glass caps are available. In

•

this case, all positions must be equipped with samples!
When operating with 300 ml sample tubes, close unused positions with a glass cap towards the

•

suction unit.

6 .7 .4

Installing 500 ml sample tubes

Unlike 300 ml (and 250 ml) sample tubes, 500 ml tubes can not simply be inserted into the appro-
priate rack. The information in this section serves to set up this type of sample tube properly and
safely in the rack.

Installation steps:

For installation and handling all parts must be < 40 °C!

*CL

ICK

*

1

2

3

frontside

rear side

Plug the rack into the carrier

•

1.

➥ Two audible *click* sounds will appear as soon as the rack is securely locked at the carrier!

➥ Check the locking by lifting the rack together with the carrier (carrier must follow the rack)