Click the name of the desired configuration to highlight it. Then click
OK. Selected settings will be restored.

The Load Instrument and Tray Settings window

miniAV-LT analysis modules

The VISCPRO

®

software is shipped with several analysis modules which

can be used to create reports based on sample test data. For general
instructions on using analyses, see Chapter 6, Analysis Configuration
Options
. For information on a specific analysis module included with
your miniAV-LT package, see the chapter on the particular analysis.

Configuring the VI Matcher

VI Matching with VISCPRO

®

The VI Matcher is a VISCPRO

®

software feature that functions in

tandem with the VI analyses to enable Viscosity Index (VI) calculation
from sample data residing in the VISCPRO

®

database.

Viscosity Index calculations require kinematic viscosity data derived
from samples tested at a high and a low temperature. The VI Matcher
permits the user to define which viscometer tubes are high temperature
tubes and which are low temperature tubes.

After the VI Matcher has been initially configured by the user (this is
ordinarily done during initial instrument setup), the VI Analysis can be
used to automatically scan the VISCPRO

®

sample database for samples

with matching Sample IDs. If samples with an identical Sample ID have
been tested on a low temperature tube and a high temperature tube within
an 8-hour time span, the VI for those samples can be calculated and