Checkline DT-326 User Manual

Page 21

Advertising
background image

18

4.4.2.2 “Delay angle” setting

Set the delay angle from 0° to 360° in 1° increment s.

The actual delay time is as follows:

While the DT-326 does not flash at the 1

st

trigger pulse as shown below in the timing chart.

[Example]

Trigger: Positive edge

Delay angle: 36°

If the frequency of the external trigger input changes, the timing of the flash is inaccurate, because the timing is

calculated based on the previous measured period.

If the current period of external trigger input is less than the previous period and the next trigger input occurs

before the flash time, the delay angle setting is ignored and the unit flashes at delay angle = 0°.

Delay angle increases as the dial is rotated to the right. The angle settings will go to 0° when angle is past 359°.

Operation

Expression






Turn clockwise

Auto Range
Trigger
Edge

Delay Deg

Flash Time

Ave Last Min

Max

min

sec

h

Hz

ms

Auto Range
Trigger
Edge

Delay Deg

Flash Time

Ave Last Min

Max

min

sec

h

Hz

ms

Auto Range
Trigger
Edge

Delay Deg

Flash Time

Ave Last Min

Max

min

sec

h

Hz

ms

Go to delay time setting

Delay angle setting

360°

× Period of External Input + approx. 60us (Built-in)

PHASE+

External Pulse
Input

LED Flash

100ms (360°)

10ms (36°)

Delay

80ms (360°)

8ms (36°)

Delay

Measure

frequency

Measure

frequency

Advertising