Model 5i Digital Force / Torque Indicator

User’s Guide


21

POL

Mitutoyo output with polarity (+ for compression, - for tension)

NPOL

Mitutoyo output without polarity (absolute value)

PM

Print/send data to a Mitutoyo-compatible device

Averaging

A

Enable Average mode

AD

Disable Average mode

AM

Select Average mode (if enabled) for primary reading

ATn

Average time. n=0.1-300.0 seconds

DELn

Initial delay. n=0.1-300.0 seconds

TRFn

Trigger force. n=value (+ for compression/clockwise, - for tension/counter-
clockwise)


External

Trigger

ETH

Enable high level-triggered External trigger mode

ETL

Enable low level-triggered External trigger mode

ETHL

Enable reading captured on a high to low transition

ETLH

Enable reading captured on a low to high transition

ETD

Disable External trigger mode


Input / Output Bits

Sn

Set output bit (open drain, pull to ground). n=0,1,2

Cn

Clear output bit. n=0,1,2

Rn

Read current status of output bit or level of input pin. n=0,1,2,3

Personality

RN

Read product name

RM

Read

model

number

RV

Read firmware version number

RS

Read serial number

Other

Commands

AOFFn

Auto-shutoff. n=0-30 minutes. 0=auto shutoff disabled

SAVE

Save current settings in nonvolatile memory

LIST

List current settings and status


Following is an example LIST output: V1.00;LBF;CUR;FLTC8;FLTP1;AOUT00;AOFF5;FULL;MIT;POL;B0

All fields are separated by “;”. The first field shows the firmware version, the last field shows the remaining
battery power (B0=full charge, B3=minimum power). All other fields show the status of settings and
features using the same abbreviations as the commands to set them.

Any detected errors are reported back by means of the following error codes:

*10

Illegal

command

*11

Not

applicable

*12

Invalid

specifier

*22 Value too large


10 CALIBRATION

10.1 Initial Physical Setup