Model 5i Digital Force / Torque Indicator

User’s Guide


25

12. Any errors are reported by the following screens:

Displayed at the start of calibration if a disallowed unit is selected.

Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.

The calibration weight does not match the set value.

The entered calibration point is too close to the previous point.

11 PASSWORDS


Two separate passwords may be set to control access to the Calibration section and to the menu and
other keys. To access the passwords setup screen, select Passwords from the menu. The display will
appear as follows:

PASSWORDS

Calibration

Menu Key
Units Key
Mode Key
Zero Key
Data Key

CALIBRATION

TENSION

LOAD TOO CLOSE

TO PREVIOUS

PLEASE TRY AGAIN


LOAD NOT STABLE

PLEASE TRY AGAIN

CALIBRATION

Units must be gF.

PLEASE TRY AGAIN

PRESS ENTER

CALIBRATION

COMPRESSION

LOAD TOO LOW

PLEASE TRY AGAIN