Using the Control Unit

UP

Press and hold for tension, release button to stop motion. Located

on the face of the Control Unit.


DOWN

Press and hold for compression, release button to stop motion. Lo-

cated on the face of the Control Unit.


SPEED CONTROL DIAL
Adjust speed by turning the dial 0.5 – 13 in/min [13 – 330 mm/min].

Located on the face of the Control Unit.


POWER SWITCH
Use this switch, in the left rear of the Control Unit, to turn on and

turn off power to the test stand. Power is indicated by an amber light
on the face of the Control Unit


POWER PLUG RECEPTACLE
Located in the left rear of the Control Unit. Plug the power cord in

here.


CONTROL CABLE
Plug this cable into the lower male connector on the test stand, ad-

jacent to the motor.

Models ESM & ES30

5

ESM

SPECIFICATIONS

Load capacity

200 lbs [1000 N]

Speed range

0.5-13 in/min [13-330 mm/min]

Maximum travel

13.0 in [228.6 mm]

Maximum travel w/options

11.75 in [298.45 mm]

Speed variation with load

±0% (Stepper motor driven)

Speed accuracy

±5% of setting

Overload protection

Max. 150% of full capacity

Power

Universal input 80-240 VAC, 50/60 Hz

Fuse type

1.2 A, 250V, 3AG SLO BLO

Weight (test stand, only)

20 lbs [9.1 kg]

Control unit weight

2.7 lbs [1.2 kg]