Checkline HTG2 User Manual

Page 6

Advertising
background image

page 11

page 6

RECHARGING NI-MH BATTERY

1. To maximize the life of the battery, power is automatically shut

off after 10 minutes of non-use or user-defined interval.
Automatic shut off is bypassed during USB output or when
used with the AC adapter/charger.

2. Battery icon will flash when the gauge needs to be

recharged.

3. Push

to turn off power. Only use the IMADA AC adapter/

charger provided, AD120 for 115VAC, AD230 for 230VAC. Plug
into the correct AC output. It takes 10 hours to charge fully.

4. When the gauge is turned off, make sure the AC adapter/charger

is disconnected to avoid overcharging.

OPTIONAL CABLES

10' Analog cable
CB-104

10' RS-232C cable, 9 pin female
CB-204

10' Digimatic Cable
CB-304

Storing Data into Memory

During measurement whether Peak or Real Time, press

to store

and display up to 1000 torque values into memory. (If no data is

stored – – – – – is displayed then flashing CF9 with solid End).

Peak or Real time Measuring Mode

Press

to turn on and the gauge automatically

enters real time measuring mode. For peak mea-

surement press

. The “Peak icon” appears on

the display. Peak readings will not change until a

higher value is measured. Press

again to

return to real time mode.
“Or PEAK“ is the factory default which measures peak CW torque
or peak CCW torque. “And PEAK“ measures both peak CW torque
and peak CCW torque during a test. Refer to the F2 function of
the Power-On programming table for the “And PEAK” function.

Tare

Clamp test sample firmly by tightening the

key. If necessary, press

to tare before the

test. Pressing

also clears the peak reading.

If High and Low setpoints have been programmed (see page 5),
for example, 5 lbf-in is set as Low and 10 lbf-in as High, the
ORANGE LED light for measurements less than 5 lbf-in (Low set-
point). GREEN lights between 5–10 lbf-in and RED lights over 10
lbf-in (High setpoint). Setpoint output is available through the
Communications port (see page 7).

After measuring, press the

button to transmit data to:

RS-232 models:

RS-232C or Digimatic devices

REAL TIME MODE DISPLAY

PEAK MODE DISPLAY

KEY

TURN MEASURING
CHUCK BY HAND

Advertising