Cal settings menu – Checkline PT-RTR User Manual

Page 13

Advertising
background image

10

Tape Grade

Cal Settings Menu

Cal Settings Menu

The probe should be zeroed regularly to ensure best accuracy.
While the user will be required to zero the probe every time the
Gage is powered-up. The Zero menu item allows the procedure to
be performed at more regular intervals. It is particularly useful
during long measurement sessions.

Clean the probe with the included cleaning card (pg. 7) before
performing a probe zero.

1. Select Zero from the Cal Settings menu
2. Simultaneously press and hold both probe buttons firmly until
the Gage beeps and the arrows point outward. Do not place
anything in the measurement opening during this procedure.

Zero

Selects a replica tape grade. Required when Linearize mode (H

L

)

is ON.

C - Course

XC - Extra Course (default)

Simultaneously press and hold both probe buttons

firmly until the Gage beeps and displays “0”. This can be
performed from the main measurement screen without having to
access the menu.

Coarse Minus and X-Coarse Plus grades are not

supported in Linearize mode since the only function of these two
grades is to improve upon the accuracy of C tape at its low end and
XC tape at its high end, something the PosiTector RTR will
automatically do when measuring H

L

. When Linearize mode is

OFF, the PosiTector RTR will measure all four grades of replica
tape just like a conventional spring micrometer.

NOTE:

SHORTCUT: