DT220 Users Manual version 1.a Configuration Parameters

18

Scale Parameter 8: Power-On Zero (N [disabled])

The power-on zero function zero’s the scale when power is turned on to the DT220.
Position the pointer to selection 8 and press the

Enter switch to toggle the Power-On

Zero function on (Y) or off (N).

Calibrate

There are 5 calibration selections.

1 Deadload first

2 Span first

3 Deadload only

4 Multi-point calib

5 Adjust Gain (tweek)

Use the

Review (↑↑↑↑) and Cont. (↓↓↓↓) keys or the numeric keypad to move the pointer to a

menu item and then press the

Enter key.

NOTE: Set the Scale Parameters and save into memory before calibrating the scale.

Calibrate 1: Deadload first

This function calibrates the scale by reading the deadload first and then reading the span.

Best results are obtained when the DT220 and load cells have reached thermal

equilibrium (power applied for approximately 30 to 120 minutes depending on load cell

size).
1. The DT220 prompts “Press ENTER to read deadload”. Remove all weight from the

scale. Press the

Enter key. The DT220 reads the scale’s deadload weight. NOTE:

press the

Cont. key to skip deadload read.

2. The DT220 display prompts “Enter calibration weight:”. Place the calibration

weight on the scale. Enter the calibration weight using the numeric keypad. Press

the

Enter key. The DT220 reads the calibration weight.

Calibrate 2: Span First

This function calibrates the scale by reading the span first and then reading the deadload.

Best results are obtained when the DT220 and load cells have reached thermal

equilibrium (power applied for approximately 30 to 120 minutes depending on load cell

size). The deadload must be greater than approx. 1% of scale capacity. If it is not, then

calibrate deadload first.
1. The DT220 display prompts “Enter calibration weight:”. Place the calibration

weight on the scale. Enter the calibration weight using the numeric keypad. Press

the

Enter key. The DT220 reads the calibration weight.