19. SPECIFICATIONS

FZ-120

i

WP FZ-200

i

WP FZ-300

i

WP FZ-1200

i

WP FZ-2000

i

WP FZ-3000

i

WP

Weighing capacity

122

g 220

g 320

g 1220

g 2200

g 3200

g

Maximum display

122.084 g 220.084 g 320.084 g

1220.84 g

2200.84 g

3200.84 g

Minimum weighing value (1 digit)

0.001 g

0.01 g

Repeatability (Standard deviation)

0.001 g

0.01 g

Linearity

±0.002 g

±0.02 g

Stabilization time (typical at FAST)

Approx. 1 second

Sensitivity drift
(10°C to 30°C / 50

°F to 86°F)

±2 ppm/°C

±0.010 g

±0.10 g

±0.15 g

Accuracy right after calibration using the
internal mass. See notes below

*

The value above is to the weighing capacity.

Clock function

Buit-in clock

Operating environment

5°C to 40°C (41°F to 104°F), 85%RH or less (No condensation)

Display refresh rate

5 times/second, 10 times/second or 20 times/second

Minimum unit mass

0.001 g

0.01 g

Counting
mode

Number of samples

5, 10, 25, 50 or 100 pieces

Minimum 100% reference mass

0.100 g

1.00 g

Percent
mode

Minimum 100% display

0.01%, 0.1%, 1% (Depends on the reference mass stored.)

External calibration weight

100 g

50 g

200 g
100 g

300 g
200 g
100 g

1000 g

500 g

2000 g
1000 g

3000 g
2000 g
1000 g

Weighing pan diameter

130 mm

150 mm

Net weight

Approx. 3.2 kg

External dimensions

193 (W) x 262.5 (D) x 84.5 (H) mm

AC adapter

Confirm that the adapter type is correct for the local voltage and power
receptacle type

Power consumption

Approx. 11VA (supplied to the AC adapter)

Interface

RS-232C

Dust and water protection

Complying with IP65

*

Accuracy right after calibration using the accessory small breeze break and the internal mass,
in good ambient conditions (within the temperature range of 10

°C to 30°C (50°F to 86°F) with

no abrupt changes in temperature or humidity, no drafts, no vibration and no effect by
magnetic fields or static electricity).

With the FZ-iWP series balance, the environment and passing age may cause internal mass
value error. Calibrate using an external mass and perform maintenance.