ALTANA BYK-mac COLOR Short Instructions User Manual

English, Operation, How to calibrate

Advertising
background image

269 020 584 E 0806

Additives & Instruments

A member of

Measure what you see.

BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments

Please note detailed operating instructions in English on CD-ROM.

ENGLISH

* Only available with BYK-mac

Operation

mode wheel:

•

Switch on

press

•

Show menu

press

•

Select function

scroll

•

Activate function press

esc button:

•

Change between menu

and measurement mode

std button:

•

Short-cut to measure standard

Operate button:

•

Switch on, select and measure

How to Calibrate?

Black/white calibration

•

Select Setup – Calibrate – Black/White Calibration

•

Place unit on black trap and confirm OK

•

Place unit on white standard and confirm OK

Black/white calibration is requested every 3 months.

Cyan check
After a black and white calibration the instrument must be checked on the
Cyan standard.

•

Place unit on Cyan standard and confirm OK

The Cyan check is requested every 7 days.

operate

How to measure Differences?

•

Select a stored standard, use

„AUTO standard“ or measure a standard

•

Input standard name:

Change character ==> turn scroll wheel
Move to next digit ==> press scroll wheel
Move one digit back ==> press esc

•

Input number of readings

•

Place instrument on standard

•

Press operate to take a reading

•

Copy predefined settings for data display

•

Continue with sample readings

Note: All color data inclusive spectral data are
saved and transferred to software.

•

First page displays color

differences

•

Scroll to second page for sparkle

and graininess results*