ALTANA Spectro-Guide User Manual

English, Spectro-guide operation, How to calibrate

Advertising
background image

269 020 584 E 0806

Additives & Instruments

A member of

Measure what you see.

BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments

Please note detailed operating instructions in English on CD-ROM.

ENGLISH

spectro-guide

Operation

Cursor buttons:

•

Navigate through main menu

and open submenu

std button:

•

Measure standard

Operate button:

•

Switch on and select

•

Measure sample

How to Calibrate?

Black/white and gloss calibration

•

Select Option – Calibrate - Calibrate

•

Place unit on light trap and press operate

•

Place unit on white standard and press operate

•

Place unit on high gloss standard and press operate

Green check
To audit instrument performance and the condition of the calibration tiles,
a check on the green standard is recommended every 7 days.

•

Select Option – Calibrate – Check Green

•

Place unit on green standard and press operate

After the green check is completed, the instrument tells you whether the
black/white calibration is still valid.

How to measure Differences?

•

Make sure to be in the main menu

•

Place instrument on standard

•

Press std button to take standard reading

•

Place instrument on sample

•

Press operate button to take sample

reading

•

Displays shows absolute sample values

and difference data

==> can be changed in Display -

Difference

Note: Readings are saved automatically when Auto Store is activated in
Option – Setup.

How to save Measurements?

•

To save standards select File – Save –

Standard

•

Input standard name:

change character ==> cursor up/down

move to next digit ==> cursor right

move one digit back ==> cursor left

Confirm ==> operate button

•

Input tolerances:

follow the same principle

•

To save samples select File – Save – Sample