Care and Maintenance

Storage: After measurement, wrap cuff around gauge and bulb and store in zippered carrying case.

Cleaning: Digital Manometer: Your ADC

®

DIAGNOSTIX

TM

brand e-sphyg

TM

gauge requires minimal

care and maintenance.

The manometer may be cleaned with a slightly dampened soft cloth, but should not be dismantled

under any circumstances.

Cuff: Sponge with a damp cloth or alcohol pad. If necessary, it can be washed in cool water with

mild soap or detergent-disinfectant. Remember to remove bladder from cuff prior to washing. After

washing, it should be rinsed and allowed to air dry. NEVER IRON. Be sure to remove the bladder

and all components before cleaning the cuff.

Bladder and Bulb: Remove bladder from cuff and disconnect bulb from tubing and valves. Wipe

with a damp cloth. To help preserve it from cracking or discoloring, wipe with a soft cloth moistened

with ethanol.

Valves: Clean the surface with a dry fabric such as soft cotton material. Remove dust from filter

screen.

Carry Case: Wipe with a damp cloth and dry thoroughly with a dry cloth. If

necessary, it can be washed with cool water and mild soap. Do not allow to air dry.

Batteries: If the e-sphyg

TM

will not be used for an extended period of time, remove batteries.

Disinfection and Sterilization

Disinfecting: Glutaraldehyde-type liquid disinfectant may only be used on the cuff. Follow instruc-

tions for use provided with the glutaraldehyde product. Prolonged use of this disinfectant may

cause discoloration. DO NOT use glutaraldehyde-type liquid disinfectant on the digital gauge, bulb,

valves, or bladder.

Sterilizing: The blood pressure cuff may be gas sterilized. DO NOT use steam or heat to sterilize the

cuff, bulb, valve, or bladder. DO NOT attempt to sterilize the digital gauge.