Specifications, Standards – American Diagnostic Corporation (ADC) e-sphyg 7002 User Manual

Page 13

Advertising
background image

Specifications

Power Source: Two 1.5V DC LR03 (AAA) Alkaline Batteries

MEASUREMENT RANGE

Pressure: 0-299 mmHg

Pulse: 30-240 beats/min Accuracy

Pressure: ±3 mmHg

Pulse: Within ±5% of reading

Pressure Sensor: Semi-conductor

Inflation: Manual

Deflation: Manual Air Release Valve

Auto-shut-off: 1 min. after last key operation

Backlight auto-shut-off: 15 sec. after last key operation

OPERATION ENVIRONMENT

Temperature: 32~115°F (0~46°C)

Humidity: 85% Relative Humidity max

STORAGE ENVIRONMENT

Temperature: -4~131°F (-20~55°C)

Humidity: 90% Relative Humidity max (non condensing)

Dimensions: 3”(L) X 2

3/8

”(W) X 1

1/8

”(H)

Weight: 3.14 oz. (G.W.) (w/o Batteries) Type BF
Battery Life: 1500 readings with back light on

3000 readings with back light off

Device and cuff are designed to provide special protection against electrical shocks.

Specifications are subject to change without notice.

Standards

ADC’s digital aneroid sphygmomanometer is accurate to +/- 3 mmHg

and conforms to applicable sections of the following international standards:

EN /ISO 81060-1: 2012, Non-invasive sphygmomanometers - Part 1:

General Requirements

EN /ISO 81060-1: 2012, Non-invasive sphygmomanometers - Part 3:

supplementary requirements for electro-mechanical blood pressure measuring systems

ANSI/AAMI/ISO 81060-1: 2007, Manual, Electronic, or Automated Sphygmomanometers

Symbol

Definition

Important Warning/Caution


Latex-Free

Circumference Size

Conforms to EU Standards

Symbol

Definition

Authorized European

Represenative’s Information

Manufacturer’s Information

Temperature Limit

Humidity Limitation

°

F