5. Wiring

A CAUTION

■ Do not work on piping and wiring

■ Ground the LF470 properly.

with wet hands.

Wet hands can cause

O

Operating this product

system failure.

without a grounding can

cause

system

DON’T

DO

malfunction.

■ Do not modify or disassemble the

LF470 unnecessarily.

Modifying or

disassembling this product

1 can cause

electric

shock,

malfunction of or dam-

DON’T , ,

age to this product.

The label shown left is

placed near the terminal

/

board for power input.

— Be alert to

electric

shock.

Flowmeter accuracy may be affected by the way wiring is executed. Proceed

with wiring taking the following precautions:

(1) Select the cable runs away from electrical equipment (motors, transformers, or

radio transmitters) which causes electromagnetic or electrostatic interference.

(2) Deterioration of flowmeter circuit insulation occurs if the converter interior or

cable ends get wet or humidified. This in turn causes malfunction of flowmeter

or noise problems. Avoid a rainy day if the flowmeter is to be installed outdoors.

Even indoors, prevent water from splashing over the flowmeter. Try to finish the

wiring as quickly as possible

(3) The excitation and input signal cables transmit extremely minute signals. Pass each

of them alone and independently through a thick walled steel conduit, and loacate

it at as much a distance as possible from other great-current cabling, nor in parallel

with such cabling.

(4) The converter has a surge-absorbing barrier installed inside. Therefore, do not

conduct a withstand voltage test for the converter. To check the insulation of the

converter, use a voltage of 250 V dc or less.

(5)

After wiring, attach the terminal cover of the converter.

-

1 8

-