Defibtech DDU-120 Series User Manual

Page 16

Advertising
background image

8

DAC-530E-EN-BH

2.2.2 Pads

WARNING

Use only Defibtech disposable self-adhesive defibrillation/monitoring pads,
battery packs, and other accessories supplied by Defibtech or its authorized
distributors. Substitution of non-Defibtech approved accessories may cause
the device to perform improperly.

CAUTION

Follow all defibrillation pad label instructions. Use defibrillation pads prior to
their expiration date. Do not re-use defibrillation pads. Discard defibrillation
pads after use (in the event of suspected pad malfunction, return pads to
Defibtech for testing).

WARNING

The defibrillation pads are intended for one time use only and must be
discarded after use. Reuse can lead to potential cross infection, improper
performance of the device, inadequate delivery of therapy and/or injury to the
patient or operator.

2.2.3 Patient Analysis

WARNING

Aggressive or prolonged CPR to a patient with defibrillation pads attached can
cause damage to the pads. Replace the defibrillation pads if they become
damaged during use.

WARNING

CPR during analysis can cause incorrect or delayed diagnosis by the patient
analysis system.

WARNING

Do not place adult defibrillation pads in the anterior-posterior (front-back)
position. A shock or no shock decision may be inappropriately advised. The
DDU-120 AED requires that the adult defibrillation pads be placed in the anterior-
anterior (front-front) position.

WARNING

Some very low amplitude or low frequency rhythms may not be interpreted
as shockable VF rhythms. Also some VT rhythms may not be interpreted as
shockable rhythms.

WARNING

Handling or transporting the patient during ECG analysis can cause incorrect or
delayed diagnosis, especially if very low amplitude or low frequency rhythms
are present. During analysis and from “Shock Advised” until “Shock Delivered,”
patient movement and vibration must be minimized.

Advertising