8 technical specifications, 1 defibtech ddu-120 aed – Defibtech DDU-120 Series User Manual

Page 53

Advertising
background image

45

DAC-530E-EN-BH

8 Technical Specifications

8.1 Defibtech DDU-120 AED

8.1.1 General

Category

Specification

Size

8.5 x 11.8 x 2.7 inches (22 x 30 x 7 cm)

Weight

Approximately 4.2 lbs (1.9 kg) with DBP-1400 Battery pack

Approximately 4.4 lbs (2 kg) with DBP-2800 Battery pack

Equipment Classification

(per EN 60601-1)

Internally powered with defibrillator-proof BF-type patient applied parts.

8.1.2 Environmental

Category

Specification

Operating / Maintenance

Temperature

0 – 50°C (32 – 122°F)

One Hour Operating

Temperature Limit

(extreme cold)*

-20°C (-4°F)

Humidity

5% – 95% (non-condensing)

Standby / Storage

Temperature

0 – 50°C (32 – 122°F)

Humidity

5% – 95% (non-condensing)

Altitude

-150 to 4500 meters (-500 to 15,000 feet)

per MIL-STD-810F 500.4 Procedure II

Shock / Drop Abuse Tolerance

MIL-STD-810F 516.5 Procedure IV (1 meter, any edge,

corner, or surface, in standby mode)

Vibration

MIL-STD-810F 514.5 Category 20

RTCA/DO-160D, Section 8.8.2, Cat R, Zone 2, Curve G

(Helicopter)

RTCA/DO-160D, Section 8, Cat H, Zone 2, Curves B&R

(Jet Aircraft)

Sealing

IEC 60529 class IP54; Dust Protected, Splash Proof,

(Battery pack installed)

ESD

EN 61000-4-2:1998 Severity Level 4

(Open air discharges up to 8 kV or direct contact

discharges up to 6 kV)

EMC (Emission)

EN 60601-1-2:2001+A1:2006, method EN 55011:1998

Group 1 Level B

(Not to exceed 30 dB μV from 30 Hz to 230 MHz and

not to exceed 37 dB μV from 230 to 1000 MHz

EMC (Immunity)

EN 60601-1-2:2001+A1:2006, method EN 61000-4-

3:1998 Level 3

(Field strength: 10V/m; carrier frequency range:

26 MHz to 1 GHz; AM modulation, 80 percent index,

at 3 frequencies: 1, 5, and 20 Hz)

* From room temperature to temperature extreme, one hour duration.