Instruction and Operation Manual

for ATEX Approved Load Cells

Last revision: 24.10.2014

Revision: 9.5

Page 3 / 12

Date of 1st issue: 03.06.2002

2.2.1 Examples of suitable circuits with approved safety barriers :


Warning:

The displayed examples are verified for ignition protection.
The specialist who installs the equipment must take responsibility
for proper operation in combination with various measuring equipment .


Example : With safety barriers for single-ended supply

Wiring diagrams of different circuit examples are available on request

2.3 Connections of the 6-wire Version

Supply circuit:

green (+) and black (-)

Signal circuit:

white (+) and red (-)

Sense circuit:

blue (+) and brown (-)

Shield:

yellow and / or metallic


The intrinsically safe circuit including the load cells must be built up with approved safety barriers or switch
amplifiers, matching the connected weighing indicator.