C.4.2 Set up Commands – SD, MT, GA, TE, TR, TL, AI

Note: All setups should be stored with the WP command before power off.

SD

Start Delay 0 … 500 ms

Set the delay between falling/rising edge of trigger pulse and start of measurement. Permitted values are 0
… 500 ms.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

SD

¿

S+00100

Request: SD=100 ms

SD_200

¿

OK

Setting: SD=200 ms

Factory default: 0 [= 0 ms]

See time diagram of check weighing page 32

MT

Measuring Time, range 0…500 milliseconds

Set the time over which the average value will be built. Permitted values are 0 … 500 ms.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

MT

¿

M+00100

Request: MT=100 ms

MT_500

¿

OK

Setting: MT=500 ms

Note: MT = 0 means disabled trigger and average function.

Factory default: 0

See time diagram of check weighing page 32

GA

Get Average

Issuing the GA command the DAS 72.1 returns the current average weight value by using the MT setup.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

GA

¿

A+01100

Answer: GA=1100 g

Note: During the time between the trigger condition being accepted and the average value being updated,
the GA command will return the value 99 999.

TE

Trigger Edge

Issuing the TE command selects rising or falling edge trigger. Parameter = 0 select falling edge and
parameter = 1 select rising edge.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

TE

¿

E:001

Request: TE=1

TE_0

¿

OK

Setting: TE=0

Factory default: 0 [falling edge]

See time diagram of check weighing page 32