Flintec MCS-64 CANbus User Manual

Page 36

Advertising
background image

Page Manual MSC-64

36

Mode 1 Characteristic (FIR-Filter)

FL

Settling time

to 0.1%

(ms)

3 dB Cut-

off

(Hz)

20 dB

damping at

frequency

(Hz)

40 dB

damping at

frequency

(Hz)

Damping in

the

stopband

(dB)

Stopband

(Hz)

Output rate

max.

(samples/s)

0 no

filtering **

600

1 47 19.7 48 64 >90

>80

600

2 93 9.8 24 32 >90

>40

300

3 140 6.5 16 21 >90

>26 200

4 187 4.9 12 16 >90

>20 150

5 233 3.9 10 13 >90

>16 120

6 280 3.2

8 11 >90

>13 100

7 327 2.8

7 9 >90

>11 85.7

8 373 2.5

6 8 >90

>10 75

** Antialiasing filter 17 Hz @ 60 dB/decade

Attention:

In mode 1 the output rate is dependant on the selected filter level (FL) and will be

automatically adjusted by the LDM88.x.


UR

Set the update rate (average building)

[ 2100sub11 ]

This command will define over how many measurements, from the preceeding IIR or FIR

filter, an average will be calculated. The average will be calculated over 2

UR

samples.

Permitted values 0...7 (see table below).

UR

0

1

2

3

4

5

6

7

No. of samples

1

2 4 8 16 32 64 128

Check / adjustment update rate:

Master (PC / PLC) sends

Device responds

Result

UR U+0001

Request: average over 2 samples

UR_4 OK

Setting: average over 16 samples

Factory default: 0 [no average, = 600 samples/s]


Remark to Mode 1

Dependency Output Rate - averaging UR - Filter FL

UR

Output Rate samples/s

FL0

FL1

19.7 Hz

FL2

9.8 Hz

FL3

6.5 Hz

FL4

4.9 Hz

FL5

3.9 Hz

FL6

3.2 Hz

FL7

2.8 Hz

FL8

2.5 Hz

0

600 600 300 200 150 120 100 85.7 75

1 300 300 150 100

75

60

50 42.85 37.5

2 150 150 75

50 37,5 30

25 21.42 18.75

3

75 75 37.5 25 18.75 15 12.5

10.71

9.38

4 37.5 37.5 18.75 12.5 9.38 7.5 6.25 5.36

4.69

5 18.75 18.75 9.38 6.25 4.69 3.75 3.13 2.68 2.34
6 9.38 9.38 4.69 3.13 2.34 1.88 1.56 1.34 1.17
7 4.69 4.69 2.34 1.56 1.17 0.94 0.78 0.67 0.59