Junction Box Type KEK – Technical Manual, Rev. 1.00 August 2007

Page 3 of 4

I

NTRODUCTION AND TECHNICAL

D

ATA

The stainless steel junction box is designed for the parallel connection of up to 4 load cells.

Type

No. of load cells

Housing size

Inputs

Output

KEK-4

up to 4

105 x 210 x 40 mm

4x PG9

1x PG9

The junction box type KEK can be connected to the instrumentation with a shielded 6-wire signal cable.
The corner correction is done with resistors and/or potentiometer.

Housing material:

Stainless steel

Protection class:

IP65

Cable connection:

With clamping terminals

Corner correction:

By exchangeable resistors

M

ECHANICAL

I

NSTALLATION

Look for a mounting location which is more or less dry and protected from environmental stress.

E

LECTRICAL

C

ONNECTIONS

Figure 1: Dimensions in [mm]

The connection sequence of the load cells should correspond to the corners of the scale, i.e
Corner 1 = Load cell 1, Corner 2 = Load cell 2, etc.

L

OAD

C

ELL

C

ABLE

C

ONNECTION

First the cable gland (see (1) in fig.1) must be loosened. Then you have to feed the load cell cable through the
cable gland unless the shrink tube is fully disappeared in the box. The wires have to run below the printed
circuit board and will be pulled back to the top at the upper end of the printed circuit board. Afterwards you can
connect the cables to the clamping terminals as indicated below:

Cable coulour

Description

Terminal designation

yellow

= Cable shield

Shield / Schirm

red

= Signal – (Output –)

red / rot

white

= Signal + (Output +)

white /weiss

black

= Excitation – (Input –)

black / schwarz

(if applicable, brown)*

= Sense –

black / schwarz

green

= Excitation + (Input +)

green / grün

(if applicable, blue)*

= Sense +

green / grün

* if load cell is equiped with 6-wire conductor cable