Flintec FT-11D User Manual
Page 41

FT-11(D) Technical Manual, Rev. 1.35 November 2010
Page 41 of 52
Examples:
The commands will be used for an indicator which instrument address is set to 1.
Reading weight data:
@010300000002FA<Cr>
Reading status data:
@010300020001F9<Cr>
Reading Tare data:
@010300030002F7<Cr>
Taring:
@011000080001020002E2<Cr>
Loading 1500 value to Setpoint 1:
1.step:
@0110000B000204000005DCFD<Cr>
2.step:
@011000080001020006DE<Cr>
Reading Setpoint 1 (SP1=1500):
1. step:
@011000080001020007DD<Cr>
2. step:
@0103000B0002EF<Cr>
Break down an example:
Setpoint 1: Write value 1500 into buffer
@
01
10
000B
0002
04
000005DC
FD
<Cr>
STX Device-
Adresse
Write Add.
Reg.
for
Setpoint 1
No. of
Addr. Reg.
No. of
Data Reg.
Value 1500 (dec.),
shown in hex
Checksum EOP
NOTE:
The points you should paid attention to while loading and reading setpoints are:
To write any value to FT-1x the value should be loaded into the buffer (address 000B). Afterwards send
command “6” to address 0008 to make FT-1x to use the value at address 000B.
To read a value from FT-1x you should send command “7” to address 0008 to make FT-1x to load the
value into the buffer. Afterwards you can read the value at address 000B.
The commands “6” and “7” which are sent to address 0008 let you write or read all setpoints