FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

5.5.4 Diagnostics [9--]

The operations about checking and testing FT-10 can be made here.

[90-] Tests

[900 ] Key Pad testing

In this step every keys ASCII code will be shown on the display as you press the related key. By this
way you can test if all the keys are functional or not. Pressing <

> key will take you to the next

parameter.

[901 ] RS 232C Serial Interface testing

The characters in the alphabet will sequentially be transferred from RS 232C serial interface port by
pressing < Zero > key one after another. Received numerical data is seen on display.

[902 ] RS 485 Serial Interface testing

The characters in the alphabet will sequentially be transferred from RS 485 serial interface port by
pressing < Zero > key one after another. Received numerical data is seen on display.

[903 ] Parallel Inputs

[ı X Y]

To perform parallel input test, enter the number of parallel input to Y digits via < Zero > key.
X shows the logical condition of that input.

[904 ] Parallel Outputs

[o X Y]

To perform parallel output test, enter the number of parallel output to Y digits via < Zero > key.
Change the logical condition of that output via < Tare > key and X shows the logical condition of that
output.

[905 ] mV Indication

As you press < Enter > key the output voltage of the load cell will be shown on the display. This
uncalibrated value is only for test / service purposes.

[99-] Printing Parameter Values

[990]

Print All Parameters

By pressing < Enter > key the whole parameters can be printed.

[991] Load Default

Press < Enter > key. [ Ld dEf ] message appears on the display. Press < Tare > key for loading
default parameter values or press < F > key to go [ 9- ] sub block.
The scale build parameters and calibration is not changed.