Grant Instruments RC User Manual

Page 4

Advertising
background image

Recirculating Chiller

Part No 09931/Issue 8 Page 4 August 2001

1 Safety


The following symbols marked on the equipment mean:-


Caution: Read these operating instructions fully before use and pay particular attention to

sections containing this symbol

Caution: Surfaces can become hot during use.

Always observe the following safety precautions

Use only as specified by the operating instructions, or the intrinsic protection may be impaired.

After transport or storage in humid conditions, dry out the unit before connecting it to the supply
voltage. During drying out the intrinsic protection may be impaired.

Connect only to a power supply with a voltage corresponding to that on the serial number label.

Connect only to a power supply which provides a safety earth (ground) terminal.

Before moving, disconnect at the power supply socket. Do not remove the IEC connector (where
applicable).

Do not check the temperature by touch, use the temperature display or a thermometer.

To reduce the risk of eye injury during high temperature operation, use safety goggles or
spectacles.

Do not touch surfaces which become hot during high temperature operation.

Ensure that the mains switch is easily accessible during use.

Do not block or restrict ventilation slots.

If liquid is spilt inside the unit, disconnect it from the power supply and have it checked by a
competent person.

It is the users responsibilty to carry out appropriate decontamination if hazardous material is spilt
on or inside the equipment.

Do not connect to a power supply or switch on before filling the system.

To prevent excessive pressure occurring during closed loop operation fill the system to a
maximum of 95% of the total volume to allow for liquid expansion.

Take care when topping up or draining, as the liquid in the system may be very hot or cold.

• If

the

alarm lamp is illuminated do not touch the liquid, it may be very hot. Refill carefully.

Drain before moving the unit. Before draining allow the liquid to cool to below 50°C or heat to
above 10°C.

Use only liquids specified in the operating instructions, within the specified temperature range. If
the alarm lamp is illuminated the liquid temperature may be above its recommended maximum:
do not inhale the vapours given off as they may be toxic; liquids should be safely discarded and
replaced.

When pumping to external apparatus, always use connecting pipe or hose suitable for the
operating temperature and liquid used, and check that the pipe connections are secure.

Do not disconnect any pipes or hoses while they contain very hot or cold liquid.

Refrigerated circulators are heavy, take care when lifting or moving; empty weights vary from
24kg to 88kg.