Recirculating Chiller

Part No 09931/Issue 8 Page 8 August 2001


2.6

Remote Alarm Socket

This alarm is operative whenever the liquid temperature goes outside the user-setable bands
around the set point. The factory setting for these alarms is 10

°C above and below the set point;

to alter the setting, refer to section 3.3. During the initial heat up or cool down phase, or
whenever the set temperature is changed, the alarm is disabled. The alarm is also operated by
the flow fail and a separate overtemperature cutout that operates when the equipment has gone
outside its upper range. The alarm relay contacts are connected to a three pin Alarm output
socket on the rear of the unit, for the attachment of an external alarm device - a light or buzzer for
example. The contact ratings are 2A at 240V

Connect the external alarm socket pins as follows:

Pin1

Ground

Pin 2 Output

Pin 3 Output


2.7 Remote

Probe


Plug remote probe into socket on rear panel, if required.

Place remote probe in liquid which is to be temperature controlled. Clamp in position so that it
cannot be accidentally pulled out of liquid.

3 Operation



Grant

flow fail cooler

Alarm override on water / glycol power

¦

¦

¦

¦

off water












Figure 3

Power switch

Water / glycol switch

Cooler on / off switch

Flow fail override switch

General alarm light

Switch to water if water is being used.
Switch to water / glycol if water /
glycol is being used

Press to turn on cooler

Press for five seconds after switching on the
main power switch or to restart after flow fail

Lights during alarm state

Not applicable for RC3000G

Controller (see fig 4)