Troubleshooting, Control panel does not respond, Tamper condition does not restore – Interlogix TX-6310-01-1 User Manual

Page 4: If a tamper alarm occurs, Alarm/open condition does not restore, Specifications, Regulatory information

Advertising
background image

Tamper condition does not restore:

A canned CO testing agent must be used for the CO functional
gas test.

Make sure the CO alarm body is properly installed on the
mounting plate.

1. Wait at least 10 minutes after installation to test the CO

alarm.

Make sure there are no trouble indications at the CO
alarm.

2. Make sure the green Power LED is flashing for normal

operation.

Make sure that you are using a compatible control panel
(see “Specificationsbelow).

3. Set the control panel to sensor test mode.

4. Press and hold the Test/Hush button until the unit beeps

three times (approximately 10 seconds), and then release
the button. The unit will enter the functional gas test mode.
The Power LED will blink once per second while in
functional test mode.

If a tamper alarm occurs:

Make sure that you are using a compatible control panel
(see “Specificationsbelow).

Make sure that the control panel is in sensor test mode
during sensor testing.

5. Apply UL approved CO test agent to the slit as shown in

Figure 7 below. When CO is detected, the unit will activate
a CO alarm. The unit will send RF test packets to the
control panel when the CO alarm is activated.

Alarm/open condition does not restore:

Make sure that the CO alarm condition has cleared at the
CO alarm.

Figure 7: CO testing slit

Make sure that you are using a compatible control panel
(see “Specificationsbelow).

CO testing slit

Specifications

Compatible panels

Simon XT (firmware 1.4 or greater) from
UTC Fire & Security

Power

Three AA batteries

Battery type

1.5 VDC alkaline

Required batteries

Duracell MN 1500, Duracell MX1500,
Energizer E91

Sensor Electrochemical

Sensor life

7 years

Frequency 319.5

MHz

Supervisory interval

64 minutes

Audible alarm

Temporal 4

Alarm response times

70 PPM = 60-240 min.
150 PPM = 10-50 min.
400 PPM = 4-15 min.

Dimensions

4.68 x 2.75 x 1.85 in. (119 x 70 x 47 mm)

Storage temperature

-4 to 140°F (-20 to 60°C)

Operating environment
Temperature
Relative

humidity


40 to 100°F (4.4 to 37.8°C)
10 to 95% noncondensing

6. The control panel will beep and display the number of RF

packets received.

7. At the control panel, exit sensor test mode.

8. Exit functional gas test mode:

Press and release the Test/Hush button; or
A 2 minute timeout will automatically cause the CO to
return to normal operating mode.

Troubleshooting

This information is provided to help you diagnose and solve
various problems that may arise while configuring or using the
wireless CO alarm.

Unit does not power up properly or reports low battery:

Make sure the batteries are fully seated within the battery
compartment and the polarity is correct.

Make sure that all three batteries are installed.

Regulatory information

Check the battery voltage (1.5 VDC nominal per battery).

ETL rating

ANSI/UL 2034

FCC compliance

This device complies with part 15 of the FCC
Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any
interference received, including interference that
may cause undesired operation.

FCC ID: B4Z-929A-CO

Control panel does not respond:

Use the 60-401 RF Sniffer to make sure the CO alarm is
sending messages for activation.

Move or rotate the CO alarm position.

Make sure the CO alarm is properly enrolled into the
control panel.

Make sure that you are using a compatible control panel
(see “Specifications” below).

4 / 8

P/N 466-2393 • REV A • ISS 14FEB11

Advertising