• If the value of the adjustment weight needs to be changed, use the
navigation buttons to make the required setting (see chap. 7.1.1.
“Numerical input”).
• The adjustment protocol output is started upon connection to an
optional printer and activated GLP function (see chap. 6.3).
When the adjustment weight display flashes, e.g.
50g, place it carefully in the centre of the sample
dish.
Adjustment starts.
Wait until „0.000g“ appears. Remove the
adjustment weight during the flashing display.
The adjustment is completed when “END” is
displayed. The device returns to the menu
automatically.
Press ESC to return to weighing mode.