KERN & SOHN DBS 60-3 User Manual
Page 34

34
DBS-BA-e-1311
Call up menu:
Press the Menu button to access the menu and
the first menu item “PRoGRM” will be
displayed.
Use navigation buttons to select menu
item “CAL”.
Acknowledge using ENTER key, the current
setting is displayed. If necessary, use
navigation buttons
to select “TEMP”.
Press ENTER to confirm and the first
temperature point to be tested will be displayed.
Press START and the first heating-up period
will start.
• Do not leave the heated cover open for more than 1 minute during the
heating-up period, otherwise the error message “ERR.100” will appear.
In this case press ESC and restart menu item “TEMP”.
• Attention: some parts e.g. ventilation grilles and inspection windows
may become very hot during operation.
• The adjustment must be made within 15 minutes when the temperature
display flashes, otherwise the temperature adjustment will be aborted
(ABORT will be displayed).
Temperature calibration for the first point takes
15 min. Compare the displayed temperature
value on the thermometer with that of the
moisture analyser whilst the temperature
display is flashing. If the two values do not
match, correct them using the navigation
buttons
and confirm with the ENTER
button Numerical input, see chap. 7.1.1.