TD GOLD 40

TD_GOLD_40-BA-e-1210

2

of 95%. This tolerance is the value by which the measured
value may vary from the real or true value. As in this
procedure, two measured values are compared with each
other, it is recommended to use a broad tolerance.

Factory-provided is this software with a safety level of
95%. This can be overwritten by any other value.


G. Material Composition


Pure gold bars or coins are rather uncommon. Typically,
gold pieces are a composition of gold and other materials,
e.g. to provide a higher stability. That means that other
materials are added in the production process. Common
are amongst others: silver and copper.

The exact composition of the gold piece can be seen in the
exposé that typically is added with bars or coins.
Alternatively, this composition can be inquired at the
precious metals separation works, the mint or the refining
establishment.

In this window, the composition is to be set in.

Please click into the line „new component“. Here please
now select the first component of the composition – other
then gold – and define the share of mass in percent of the
composition. The software now adds automatically the
relevant mass.

After adding all components, the software shows the exact
and individual sound velocity of the test object.


This individual sound velocity is to be set into the SAUTER
TD Ultrasound instrument.

H. Setting in the instrument TD special gold

Please turn the instrument on. Then please press the VEL
key. Now a sound velocity appears, e.g. 5000 m/s. By
pressing the arrow keys ▲ and ▼the sound velocity can
the changed in 10 m/s steps. Please round the calculated
sound velocity accordingly.
By pressing VEL again, this setting is stored.

I. Measurement with the TD special gold

Please put a little bit coupling gel (ATB-US 03, can be
reordered) onto the test object. Place the sensor on the
gel, gently pressed by hand. The sensor has to be
connected with the TD display unit.

If the connection is fine, in the display can now be seen:

(())

This symbol shows the successful coupling. This

means, that the ultra sound travels regularly
through the material, is reflected and captured
with its echo.

Value (mm) This number indicates the thickness in mm
of the test object

J. Interpretation of the measured value

The measured value of the TD ultrasound instrument shall
be within the tolerances calculated by the SSG software.