Instruction Manual

TB

TB-BA-e-1212

1

DIGITAL COATING THICKNESS GAUGE


Models: TB 1000-0.1 F

TB 1000-0.1N
TB 1000-0.1FN
TB 2000-0.1F

Table of contents

1.

Features

2.

Specifications

3.

Front panel descriptions

4.

Measuring procedure

5.

OFFSET-Accur function

6.

Calibration

7.

Battery replacement

8.

Calibration foils

9.

Correct handling of coating thickness
measurement with external sensors

10. Considerations/ Trouble-shooting
11. Declaration of conformity

Annotation: It is strongly recommended to calibrate
the new instrument before the first use, as
described in paragraph 6. By doing this it
will be achieved a much better measure-
ment result right from the start.


1. Features


* The instrument meets ISO standards 2178. So it can

be used in laboratory conditions as well as
under “harsh field” conditions.

* The instrument is widely used to measure the thickness
of these nonmagnetic coatings: paint, plastic & PVC,
porcelain enamel, copper, zinc, aluminium, chrome,
layers of laquer, layers of paper etc.
These coatings may be put up on following magnetic
base grounds: steel, iron, nickle, etc.
* Wide measuring range and high resolution

2. Specifications


Display:

4 digits, 10 mm LCD


Range:

1000

μm and 2000µm

Resolution:

0.1

μm to 100 μm

1 µm over 100 µm

Accuracy:


Following accuracy is given for the models TB 1000-0.1F,
TB 1000-0.1N as well as TB 1000-0.1FN:

- Standard: 3% of the measured value or min. ± 2.5 µm
Is valid within a tolerance range of ± 100

μm around the

individually measured range, if a two-point calibration was
performed within this tolerance range.

For model TB 2000-0.1F accuracy is:

- Standard: 5% of the measured value or min. ± 2.5 µm
Is valid within a tolerance range of ± 100

μm around the

individually measured range, if a two-point calibration was
performed within this tolerance range.

- Off-Set Accur Mode: 1% of the measured value
or min. ± 1.0 µm
Is valid within ± 50

μm around the Off-Set Accur point.


All tolerances are valid after the adjustment procedure!

Power supply:

4 x 1.5V AA battery


Operating conditions:

Temperature

0 to 50 °C

Humidity

≤ 80 %


Size:

161 x 69 x 32 mm


Weight:

about 260 g (including batteries)


Accessory:

Carrying case

Operation manual

Instrument and sensor

Calibration foils

Base plate


3. Front panel description

3-1 Sensor
3-2 Display
3-3 Zero- key
3-4 Plus- key
3-5 Minus- key
3-6 Power- on/ Power- off key
3-7 µm/ mil key
3-8 Battery compartment/ cover