Instruction manual fl, Sauter gmbh – KERN & SOHN FL 2K User Manual

Page 5

Advertising
background image

Sauter GmbH

Ziegelei 1
D-72336 Balingen
E-Mail: [email protected]

Tel: +49-[0]7433- 9933-199
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com

Instruction Manual

FL

FL-BA-e-1411

5

to memory page. If you selected DELETE and press
ENTER key the gauge will delete current saved record and
return to memory page. If you selected DELETE ALL and
press ENTER key the gauge will delete all saved record
and return to memory page.

4) CALIBRATION

In this menu you can adjust the force gauge. Therefor you
need a password. It is 7780. Take care that you have the
needed weights to adjust the force gauge. Contact your
SAUTER distributor for details.

5) DIAGNOSTIC

This is used to check status of the load cell. If you suspect
that your load cell transducer has sustained an overload it
is possible to check the status of the load cell immediately.
Place the gauge horizontally on the flat level surface and
go to main menu page. Use UP and DOWN key to move
the cursor point to DIAGNOSTIC and press ENTER key
the display will show Diagnostic menu page. Press ESC to
return to main menu page.

Figure 8 Diagnostic Menu


If the % offset is between 5% - 10 % please contact your
supplier to arrange a recalibration of your gauge.
If the % offset is greater than 10% please contact your
supplier to arrange for load cell replacement.
These values are given as an indicator only

– the need for

calibration / repair may vary according to the individual
characteristics of the load cell.

6) ABOUT

In this menu you can choose the output port between USB
or RS232

The FL-series have an analog output +/- 2V,
In normal operation use analog value depend on the
reading value of TRACK, PEAKT and PEAKC mode. If
PASS-FAIL function is used analog value = 2V when
PASS, and analog value = 0V when FAIL.

7) ABOUT This shows the information of your gauge

(Firmware revision, Model, Capacity, Serial number). To
access ABOUT menu, go to main menu page and press
UP and DOWN to move the cursor point to ABOUT and
press ENTER key the display will show About menu page.
Press ESC key to return to main menu page.

Figure 9 About Menu