Clean the water tank every week by using medical disinfectant. First drain the distilled water out, using
cotton fabric with medical disinfectant to scour the tank wall, then clean it use distilled water.
Clean the sterilizer chamber every month by using alcohol.

9.2 Change fuse

fig9-4

9.Service and Maintainance

Fig9-4

1). Switch off the power.
2). Unscrew the fuse holder with a screw driver counter clockwise.
3). Pull the fuse holder out with your fingers.
4). Make sure to replace the cooked fuse with a correct fuse.
5). Put back the fuse holder and screw it on clockwise with a screw driver.

SEA Series Steam Sterilizers

Operation manual

13

9.1 how to clean the water tank.

9.1.1 Please clean the sterilizer water tank and waste water tank regularly.

1

Empty the tanks water, please refer to item 10.2 draining the water.

2

Open the tank cover board, you can use screw driver to loose the five

screws just as fig9-1 indication

3

pull up the water tank cover board by one hand,just, as fig9-2 indication.

4

Pull up the tank cover completely and you can disassembly the tank cover.

Refer fig9-3 indication.

5

Now you can start to clean the tank, use cotton to dip in some alcohol or

medical disinfectant to scour the tank wall,wash it by distilled water, then

empty the tank and and dry the tank.

6) After you cleaning the tank, assembly the tank cover and tighten the screws.

Fig9-1

Fig9-2

Fig9-3