Repair methods:

Cause

Details of cause

Repairing way

Power

a. Power cord is not connected.

b. Power voltage is not proper.

Checking power plug, replace fuse

Lid problem Lid can not be closed

Contact local dealer

PC board Parts can not connect well

Contact local dealer

Motor

problem

Over- heat, carbon brush use out

Stop using it and wait for cool, replace carbon

brush or contact local dealer.

Imbalance Carrier or tube is put improperly. Please refer operation description

6. Normal problem on centrifuge:

Status Description

Repair

Motor can not drive

Carbon brush connects badly.
Wire matched badly.
The outlet of controller is out of
order.

1. check carbon brush
2. check wiring
3. check the outlet of controller
4. replace controller

Motor stops during

operation

Carbon brush connects badly.
Wire matched badly.
The outlet of controller is out of
order.

1. check carbon brush
2. check wiring
3. check the outlet of controller
4. replace controller

Motor is over-heat

Temp. switch skips away.

After temp. is cooler then operate
it again.

Motor stops running

Temp. switch is out of order.
Wire matched badly.

1. check temp. switch or replace a

new one.

2. check wiring.

8