Rotor and accessory – LABEC M-SCEN-207 User Manual

Page 4

Advertising
background image

(

) ROTOR AND ACCESSORY

1. Centrifugal force calculation: Use its speed and radius to calculate force as following

formula.

RCF= (rpm)

2

x 0.00001118 x R

R: radius (cm)
RCF: Relative Centrifugal Force ( xg )

2. Acceptable speed and load:
Please refer to the form (Ⅱ-4) to see the max. speed each model of rotor and tube carrier

can be load. The load weight including: rotor, tube carrier, washer, samples, and so on.


3. Sterilizing and cleaning rotor:

Rotor is made of stainless steel or steel or aluminum with high density coating and highly
resistant to corrosion, but not fit to alkaline and strong acid substance.

Note: Don’t heat the rotor to a high temperature (100℃) to clean or sterilize it.

4. Rotor Specifications:

Tube size (φx l ) mm

Rotor

Capacity

(ml x pcs)

Max.

RCF

(

)

Max.

Min.

Cat. No.

RA-1508

15 x 8

3,310

16.5 x 122

15 x 90

2-5052-11

RA-1512S

15 x 12

3,350

16.5 x 122

15 x 90

2-5052-21

RA-5004

50 x 4

3,350

29.5 x 115

27.5 x 90

2-5052-12

RA-5006

50 x 6

3,350

29.5 x 115

27.5 x 90

2-5052-22

REA-1520

1.5 x 20

2,640

11 x 48

11 x 40

2-5052-13

** Tubes are not attachable. **

Before using the centrifuge, please must refer to the rotor’s max. speed setting value.