LABEC M-PL-600 User Manual

Page 11

Advertising
background image

Remarks:After 1 point calibration, only the zero potential of the

sensor is modified (Offset),the slope of the pH sensor is kept from last
calibration. Even to those electrodes which have good performances,
0.05pH measurement error could be reached during measurement.

2 points calibration

After 1 point calibration, LCD

disp

hes),

few

2 points

the

nt to

h

rator.

lays as above chart (“Cal2”flas

immerse the electrode into standard
buffer solution, shaking the vessel for
minutes, wait for second, press“CA L”,

the

instrument will recognize which buffer

it the instrument judge pH sensor

response to ending point, the instrument displays as above chart,
calibration has been accomplished, automatically enter the pH
measurement mode, the display shows below, “Cal3”flashes,

instrument reminds you 1 point
calibration is finished, do you wa
proceed 3 points calibration? If only 2
point calibration needed, press “READ
to exit calibration process, switch to pH
measurement mode. If the 3 points

apter should be followed by the ope

solution is it, display as well, wa

calibration is needed, the next c

Remarks: After 2 points calibration, the instrument will modify the

offset and slope of the electrode。PH measurement under this kind of
calibration status, the measurement precision is quite good.

3 points calibration

After 2 point calibration, LCD

disp

hes),

.18

lays as above chart (“Cal3”flas

immerse the electrode into third
standard buffer solution(e.g. PH9
Buffer), shaking the vessel for few

- 10 -