Appendix 1 Checking method of pH linear error

Input 177.48mV(pH4.00 buffer solution electrode potential simulate

value)、-178.07mV(pH10.01 buffer solution electrode potential simulate
value) into the instrument,temperature compensation sent to 25℃,
operate according to 2 points calibration;after calibration the instrument
displays “

”signal.(if type 640 please set the buffer solution set to

b1)Check the linear error of the instrument according to following table.

mV

414.11 354.95 295.80 236.64 177.48 118.32

59.16

0

pH

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

mV

-414.11 -354.95 -295.80 -236.64 -177.48 -118.32 -59.16

pH

8.00

9.00

10.00

11.00

12.00

13.00

14.00