SEND[CR]

Sends weight, difference between weight

and target, and signature line

Terminates ISO mode.

ACAL[CR]

None

Starts the automatic internal calibration

weight calibration routine.

TYPE 2 COMMANDS (balance input functions)

COMMAND RESPONSE PURPOSE

[register number]_STORE[CR]

None

Stores the current weight, at the greatest

possible resolution, in the specified

register.

[Decimal

number]_ENTER_[register

number]_STORE[CR]

None

Stores the number entered, in the

current units, as an additive constant in

the register specified.

[decimal

number]_ENTER_CONVERT_[register

number]_STORE[CR]

None

Stores the number entered, as a "last

front panel units" relative multiplier, in

the specified register.

[decimal number]_TARE[CR]

None

Adds the number entered, in current

units, to the TARE register. CLEAR

[CR] zeroes the TARE register.

TYPE 3 COMMANDS (balance output functions)

COMMAND RESPONSE PURPOSE

SEND[CR]

See the RS-232 Balance Message

Formats section..

Returns the contents of the display

register.

[register number]_RCL[CR]

REG XXX:[register contents][current

units or 'MULT.'] [CR] [LF]

Sends the contents of the specified

register in the current units if it is

additive or with the designator 'MULT.'

if it is multiplicative. An unflagged

number is sent with no designator. An

undefined register returns a ? with a

[CR][LF].

RCL_TARE[CR]

REG 091:[TARE register contents]

[current units][CR][LR]

Return the TARE register's contents.

RCL_[(CONVERT)(CAL)]

REG XXX:[contents of specified

register][MULT.][CR][LF]

Returns indicated register contents, if in

that mode. If not in that mode then a ?

with a [CR][LF] is returned.

RCL_PCS[CR] REG

XXX:[contents of pieces

register][last FP units/pc][CR][LF]

Returns pieces register contents

expressed as last front panel units/pc, if

in that mode. If not in that mode then a

? with a [CR][LF] is returned.

[register number]_RCL_TARE[CR]

REG XXX:[register contents][current

units][CR][LF]

Adds the specified register contents to

the TARE register.

RCL_XAVG[CR]

[average weight] [current units]

AVX[CR][LF]. [2 second delay].

[number of samples taken]

###[CR][LF]. [2 second delay].

[standard deviation] [ current units]

SDEV[CR][LF]. [2 second delay].

[relative standard deviation]

COV[CR][LF].

Sends the results of a standard deviation

run. The TBAR mode, if in use, is

temporarily suspended while this

process is going on.

Advertising