TROUBLESHOOTING GUIDE

NOTE:

Many problems such as the balance remaining in the turn on sequence, "weird" characters on

the display or "garbage" on the display are caused by connecting the power supply to the balance incorrectly.

The proper procedure is to connect the round DIN connector on the power supply cable to the balance then plug

the power supply into the wall outlet.

PROBLEM POSSIBLE

CAUSE POSSIBLE

SOLUTION

The display is blank

*Balance not turned on

*Press ON/OFF switch

*Power cable not plugged in

*Plug in power cable

*No power from AC outlet

*Turn on circuit breaker

*Change AC outlets

OL appears in display

*Maximum capacity exceeded

*Reduce container weight

*Weigh sample in smaller increments

*Switch to high range (dual range units)

*Shipping clamp is on

*Loosen clamp (see page 4)

UL appears in display

*Pan is not in place

*Insure pan is positioned properly

*Balance is out of range

*Press ZERO switch

*Shipping clamp is on

*Loosen clamp (see page 4)

The display is unstable

*Drafts (air currents are present)

*Install windscreen, relocate balance

*Vibrations are present

*Isolate balance from vibration, relocate

balance

Weight readings are incorrect

*Balance is out of calibration

*Recalibrate balance

*Balance is not level

*Level balance

*Display was not zeroed before weight

was placed on pan

*Press ZERO switch before weighing

*Object being weighed is touching

windscreen or balance case top

*Reposition object

*Unit of weight selection is incorrect

*Select proper weighing mode

Display stays in turn on sequence

*Pan not in place

*Place pan on balance

*Drafts or vibrations present

*Isolate or relocate balance

*New windscreen has been installed

*Raise pan height

*Shipping clamp is on

*Power supply connected to wall outlet

before balance connector

*Loosen clamp (see page 4)

*Unplug power supply from wall then

replug into wall

Err 1 appears in display

*Calibration weight is incorrect

*Use calibration weight that appears in

balance display

Err 2 appears in display

*Balance not warmed up enough for

calibration to be performed

*Plug in and turn on balance for 1 hour

Err 3 through 14 appears in display

*Service is required

*Call Scientech's Service Department at

(800) 525-0522 or (303)444-1361 or

email us at [email protected].