Hepa co2 filter – Shellab SCO2W-2 User Manual
Page 15
15
HEPA CO2 FILTER
The CO2 Incubator design features a HEPA filtration system with a patented copper housing. (Patent No.
6,333,004) As the chamber air is drawn through the filter system, airborne microbes and isolated
particulates are not only removed from the air, but destroyed as well. It is through copper oxidation that
the contaminants are actually killed. The HEPA filter is easily replaced without the need of any tools and
has an efficiency rating of 99.97% at 0.3 microns.
Install the HEPA Filter onto the duct prior to installing the duct in the unit.
WARNING: Ensure that when installing the duct to ensure that the duct is not touching the C02 sensor head
as this may cause erroneous readings.
Section
7
The Hepa filter attaches to a
fixed coupling ring with a direct
“snap-in” action. Push straight
up to the coupling ring to
connect. Pull the filter straight
down
to
disengage.
Replacement Hepa filters are
available from Sheldon Mfg., Inc.
Part number 2800517.