Shelves, Power source, Installation (continued) – Shellab SMI12 User Manual

Page 10

Advertising
background image

10 |

P a g e

INSTALLATION (CONTINUED)

S

HELVES

To install the incubator shelves perform the following:

1. Hold the shelf slide at about a 10° angle (Figure 2A).

2. Insert the hooked tab into the vertical slot at the desired height (Figure 2B).

3. Push the slide into the horizontal slots.

4. Repeat for shelf slide at the same height on the opposite side of the incubator.

5. Place a shelf on the shelf slides. Repeat this procedure for additional shelves as needed.

Figure 2

P

OWER

S

OURCE

Check the data plate for voltage and ampere requirements before making a connection. If the
requirements match your power source, plug the power cord into a grounded outlet. A 6ft, Nema 5-15P
power cord is provided with the unit. Supplied voltage must not vary more than 10% from the data
plate rating. Damage to the incubator may result if supplied voltage varies more than 10%.

These incubators are intended for a 50/60 Hz application. Use a separate circuit to prevent loss of
product due to overloading or circuit failure. The SMI12 (GI12) unit requires two power cords/sources;
each chamber operates independently.

Note: Electrical supply to the incubator must conform to all national and local electrical codes.