TROUBLESHOOTING

TEMPERATURE

Temperature too high

1/ controller set too high
2/ controller failed on

– call Customer Service

3/ wiring error

– call Customer Service

Display reads "HI" or "400"+

Probe is unplugged, is broken or wire to sensor is broken

– trace wire

from display to probe; move wire and watch display to see intermittent
problems

Chamber temp spikes over set point and
then settles to set point

Recalibrate

Temperature too low

1/ high limit set too low

– see section 4.5

2/ controller set too low
3/ unit not recovered from door opening

– wait for display to stop changing

4/ unit not recovered from power failure or being turned off

– incubators

will need 24 hours to warm up and stabilize
5/ element failure

– call Customer Service

6/ controller failure

– confirm with front panel lights that controller is calling

for heat
7/ high limit failure

– confirm with front panel lights that High Limit is

operating correctly
8/ wiring problem

– check all functions and compare wiring to owners

manual - especially around any areas recently worked on
9/ loose connection

– check shadow box for loose connections

Display reads "LO"

1/ sensor is plugged in backwards

– call Customer Service

2/ if ambient room temperature is lower than range of unit

– compare set

points and ambient temperature to rated specifications in Section 10.0

Unit will not heat over a temperature that
is below set point

1/ confirm that blower motor is moving and that amperage and voltage
match data plate

– check blower motor motion in shadow box and feel for

air movement in chamber
2/ confirm that set point is set high enough

–turn High Limit clockwise and

see if High Limit light comes on
3/ check connections to sensor
4/ check calibration

– using independent reference thermometer.

Unit will heat up at all

1/ verify that controller is asking for heat by looking for controller light

– if

controller light is not on, there is a problem with the controller
2/check amperage

– amperage should be virtually at maximum rated

(data plate) amperage
3/ do all controller functions work?
4/ is the High Limit set high enough?

– for diagnostics, should be fully

clockwise with the pilot light never on
5/ HAS THE FUSE/CIRCUIT BREAKER BLOWN?

Indicated chamber temperature unstable

1/ ±0.1 may be normal
2/ is blower motor working?

– remove top panel and verify movement of

Section

8