Mechanical, Other – Shellab SGO6E-2 User Manual

Page 12

Advertising
background image

12

1/ ±0.1 may be normal but gravity models may vary ±2.0 degrees
2/ is ambient room temperature radically changing

– either door opening

or room airflow from heaters or air conditioning ?

stabilize ambient

conditions .
3/ it may happen if exhaust stack is 100% open

adjust stack to at least

¼ closed.
4/ calibration sensitivity

call Customer Service

5/ Thermostat set too low

– be sure that it is more than five degrees over

desired set point; check if pilot is on continuously; turn controller knob
completely clockwise to see if problem solved then follow instructions in
section 6.5 for correct setting.

Will not maintain set point

1/ assure that set point is at least 5 degrees over ambient room
temperature.
2/ see if ambient is fluctuating

MECHANICAL

Door not sealing

1/ Confirm that the door gasket is aligned properly.
2/ Confirm that unit has not been damaged and that the body is square.

OTHER

Unit or wall fuse/circuit breaker is blown

1/ check wall power source
2/ compare current draw and compare to specs on data plate
3/ see what other loads are on the wall circuit

Unit will not turn on

1/ check wall power source
2/ check fuse/circuit breaker on unit or in wall
3/ check all wiring connections, esp. around the on/off switch

Unit is smoking

– out of box

Put unit under vent and run at full power for one hour.

Contamination in chamber

1/ See cleaning procedure section 7.1
2/ Develop and follow a standard operating procedure for specific
application; include definition of cleaning technique and maintenance
schedule

Advertising
This manual is related to the following products: