7

CONTROL PANEL OVERVIEW

Figure 2

4.1

Power Switch: The main power I/O (on/off) switch controls all power to the oven. It
must be in the I, or ON position before any systems are operational.

4.2

Main Temperature Controller: Marked SET TEMPERATURE, this control is
equipped with an adjustment knob and a graduated dial. The graduated dial is
marked with 10 major increments and 50 minor increments. The increments can be
used as index points for setting and returning to set point temperatures.

4.3

HEATING light: This green pilot light is marked HEAT ACTIVATED and indicates
when the element has been activated and the oven is heating. When set point is
reached the pilot light will cycle on and off as the elements maintain the temperature
selected.

4.4

Over Temperature Thermostat: Marked SET OVERTEMPERATURE, this control
is completely independent of the Main Temperature control and is equipped with an
adjustment knob that requires a flat-edged tool when making adjustments to
eliminate accidental changes. It provides safety temperature protection for the oven.

If for any reason the oven temperature rises above the Main Temperature control’s

set point, the Over Temperature Thermostat will limit the rise to approximately 10

C

above the set point selected.

4.5

Over Temperature light: This red pilot light is marked OVERTEMPERATURE
ACTIVATED and is on when the Over Temperature Thermostat has been activated
and taken control of the elements. Under normal operating conditions this pilot light
should never be on.