Troubleshooting, Temperature, Refrigeration – Shellab SRI3-2 User Manual

Page 9

Advertising
background image

9

TROUBLESHOOTING

FOR PERSONAL SAFETY, ALWAYS DISCONNECT THE POWER BEFORE SERVICING.

In the event the incubator does not operate properly, check the following before calling for service.
Always make a visual inspection of the incubator and control console when troubleshooting to find loose
or disconnected wires which may be the source of trouble.

TEMPERATURE

Temperature too high

Controller set too high.

Display reads "HI"

Probe is unplugged, is broken or wire to sensor is broken.

Temperature too low

1/ Overtemperature thermostat set too low.
2/ Controller set too low.
3/ Unit not recovered from door opening

– wait for display to stop

changing.

4/ Unit not recovered from power failure or being turned off

– incubators

will need 24 hours to warm up and stabilize.

Display reads "LO"

Sensor is shorted.

Unit will not heat up at all

1/ Verify that controller is asking for heat by looking for controller light
2/ Do all controller functions work?
3/ Is the Overtemperature set high enough?

Indicated chamber temperature unstable

1/ ±0.1 may be normal
2/ Is fan working?
3/ Is ambient room temperature radically changing

– either door opening

or room airflow from heaters or air conditioning ?

– stabilize ambient

conditions

4/ Overtemperature set too low.
5/ Electrical noise

– remove nearby sources of RFI including motors,

arcing relays or radio transmitters

Display and ref

erence thermometer don’t

match

Calibration error.

Can't adjust set points or calibration

1/ Turn entire unit off and on to reset
2/ If repeatedly happens, call Customer Service

Calibrated at one temperature, but not at
another

This can be a normal condition when operating temperature varies widely.
For maximum accuracy, calibration should be done as close to the set
point temperature.

REFRIGERATION

Unit will not cool

If the compressor is running:
1/ See if condenser is cold but free of ice
2/ Be sure that fans are circulating air: in the chamber, and over

the compressor

Section

6