Soniclean Digital Benchtop Ultrasonic Cleaners User Manual
Page 7
XX
O
C
Water temperature updated every 5 sec.
XX
100.P
80.P
60.P
100% Power levels.
Degas
dGAS
Degas displays every 5 sec.
MM.SS
Cleaning time. The display is in
MM.SS
where
MM
in minutes and
SS
in seconds. Degas cycle is: H,3-M,1-L,1-OFF,1 SEC
Power Level
100.P
80.P
60.P
Press repeatedly to scroll through and select Power levels when
unit is in standby or operating
Heater On Off
H
.
On
Heater On. Key valid when heater is enabled.
The LED will blink
when the set temperature is reached. With out ultrasonics the
H.On
display will flash after one minute to indicate that the
heater is on.
H
.
OFF
Heater OFF. Key valid when Heater is disabled.
XX
.
C
Use More or Less to set temperature. Max 60
o
C.
XX
.
C
LED lights
up when heater is active.
Status
XX
O
C
Water temperature updated every 5 seconds.
XX
O
.
C
Heater temperature set. Dot before C indicates heater set
temperature.
MM.SS
Cleaning time. The display format is
MM.SS
where
MM
is minutes
and
SS
is seconds. Total cleaning time includes degassing time.
100
.
P,
80
.
P, 60
.
P
Power level set. Display
100
.
P, 80
.
P, 60
.
P
.
XX
.
ct
Cleaning set time where
XX
is from 1 to 99 minutes.
XX
.
dG
Degas time set where
XX
is from 1 to 99 minutes.
XX
.
dc
Duty cycle factory default set is 50% showing
50
.
dc
.
Use
More or
Less key to change the values. When activated in
Setup
this
functionality will be activated when the bath temperature is close to
the set temperature.
Sxxx
Software Version
Setup
BEEP
b
.
OFF
b
.
On
Enable or disable audible key press.
Press Less key to silent operation using
b
.
OFF
Press More key
b
.
On
.
F
.
SEt
F.On
Load factory values. The display shows
F
.
SEt
.
Press More or Less key
to
toggle
F.On
The factory values are:
Ultrasonic power: 100%.
Clean Time: 15.0 minutes.
Degas Time: 5.0 minutes.
Issue date: 190913-1